DELTA 15/30 Operator Manual 
Page 2.5 
Version B.2 
 
SAFETY CLASSIFICATIONS, HAZARDS AND PRECAUTIONS 
 
 
This product  is classified as a Class IV laser product  in compliance  with FDA 21 
CFR 1040.10 and 1040.11, UL 60601-1,  EN  60601-1, EN 60601-1-2,  EN  60601-
2-22 and EN 60825-1. 
This product conforms to the requirements of Council Directive 93/42/EEC of the 
Council  of  European  Communities  (Medical  Devices  Directive).  Affixing  the  ‘CE 
Mark’ to the instrument indicates conformity to this directive.      
 
0086 
 
 
The local Laser Safety Officer should review all procedures for safety prior to system use. 
 
 
A Class IV Laser is hazardous to the eye from the direct beam and diffuse reflections. It 
also presents significant skin and fire hazard. 
 
 
Avoid eye or skin (except specific treatment) exposure to direct or scattered radiation. 
Take all necessary protective measures, as explained in the rest of this section, in areas 
where the laser is being used. 
 
 
All personnel must wear approved protective glasses appropriate to the wavelength of the 
DIOMED DELTA 15/30
 to reduce the risk of eye damage. 
 
 
The aiming beam is a Class IIIa Laser and an unprotected eye may view the beam 
scattered from a non-reflective surface. Do not stare into the aiming beam or view it 
directly with optical instruments. 
 
 
Avoid directing the laser beam anywhere other than the treatment area or calibration 
ports. 
 
Before using a fiber, check it carefully for any signs of damage during storage or transit. 
Protective caps should be in place over SMA connectors. All screws and ports should be 
secure. Do not use if there is any sign of damage.