Boston Scientific Corporation CRM312004 사용자 설명서
25
Patient
–
PRM
–
Recorder Connection
To display a tracing on the PRM and an external strip chart recorder using
the patient ECG cables, set up equipment as shown in Figure 13.
the patient ECG cables, set up equipment as shown in Figure 13.
Figure 13. PRM receives surface signal via the patient cable and then sends
surface and telemetered channels to the external strip chart recorder.
surface and telemetered channels to the external strip chart recorder.
1.
Connect the external recorder cable from the PRM to the external
recorder input ports.
recorder input ports.
•
Channel 1 (red) for a surface trace
•
Channel 2 (orange) for telemetered signal 1
•
Channel 3 (green) for telemetered signal 2
2.
Connect the patient ECG cable to the patient electrodes.
3.
Verify proper telemetry wand position.
Setup is now complete.
Telemetry Wand
Patient ECG
Cable
Cable
External Recorder
External Recorder
Cable
PATIENT > PRM > RECORDER
Channel 1 (red)
PRM
Connectors
ECG
IN
IN
Wand
IN
IN
Analog
OUT
OUT
Surface
IN
Channels 2 and 3
(orange and green)
Patient ECG
Cable
Cable