Parallel Connection
To display traces on both PRM and ECG recorders using two different pa-
tient ECG leads, set up the equipment as shown in Figure 14.
Figure 14. Parallel connection of PRM surface leads and external surface 
monitor.
1.
Connect the external recorder patient ECG leads to the appropriate 
limb electrodes.
2.
Connect the patient ECG cable to the patient electrode.
3.
Connect the external recorder cable from the PRM to the external re-
corder.
NOTE:  This configuration will add the most noise to the tracings of the 
two PRM
recorder configurations.
4.
Adjust gain and filters on the external recorder.
Setup is now complete.
NOTE:  The right leg electrode conductor on the patient ECG cable is a 
driven ground. When connecting the PRM and an external ECG monitor 
in parallel, the driven grounds for the two instruments must be connected 
to the same patient limb (e.g. right leg electrode); otherwise, noise prob-
lems will occur.
Telemetry Wand
Patient ECG 
Cable
External Recorder
External Recorder 
Cable
PRM
Connectors
ECG
 IN
Wand
IN
Analog
OUT
Surface 
IN
Patient ECG 
Cable
PATIENT > RECORDER
PATIENT > PRM
Channels 2 and 3 
(orange and green)