ZM-540PA, ZM-541PA  Operator’s Manual
WARNING
When not monitoring SpO
2
, disconnect the SpO
2
 
cable from the transmitter. Otherwise, noise from 
the probe sensor may interfere and incorrect data 
is displayed on the screen.
WARNING
•  When using the TL-201T finger probe, do not 
fasten the probe and cable to the finger by 
wrapping with tape. This may cause burn, 
congestion or pressure necrosis from poor blood 
circulation.
•  When using probes other than the TL-201T 
finger probe, to avoid poor circulation, do not 
wrap the tape too tight. Check the blood 
circulation condition by observing the skin color 
and congestion at the skin peripheral to the 
probe attachment site. Even for short-term 
monitoring, there may be burn or pressure 
necrosis from poor blood circulation, especially 
on neonates or low birth weight infants whose 
skin is delicate. Accurate measurement cannot 
be performed on a site with poor peripheral 
circulation.