Huntleigh Healthcare Limited Huntleigh Diagnostic Products Division SF1US 사용자 설명서

다운로드
페이지 48
 33
Tr
ouble Shooting
8.3  Changing the RF-channel
Wireless transmission can be disturbed by other transmitters running on the 
same frequency. If there are several Freedom units in one ward, all receivers 
must be set to different channel numbers. The operating channel is set 
using the controls located on the rear panel of the Freedom receiver. This is 
performed as follows:
 
1.  Disconnect the receiver from the mains and wait until all LEDs are off.
10
1
2.  Use the code-switches on the rear panel of the receiver to select the new  
    channel number.
   
WMTS (USA) : 
00 - 99 (100 channels - see Appendix 1) 
   
ISM :  
 
00 - 25 (26 channels - see Appendix 2)
   
3.  Re-connect the receiver unit to mains and wait until the LED test is fi nished.
4.  Insert the US-transducer unit into a charging bay and wait until charging is  
 indicated.
Now the new channel is installed for receiver and transducer. This procedure 
takes about 20 seconds; please do not interrupt this process. In case of unusual 
behaviour of this process please retry this procedure.
Note:  
Transducer frequency is set by the reciever and cannot be  
  
adjusted.
Note :  
The serial numbers from installed and matched receiver and  
   
transducer units should be recorded for future use if required.  
   
This may be important if several telemetry systems are used  
   
within a ward or confi ned area.