Huntleigh Healthcare Limited Huntleigh Diagnostic Products Division SF1US 사용자 설명서
34
Speci
fi ca
tions
9. Specifi cations
9.1 Equipment
Classifi cation
Type of protection
against electric shock.
against electric shock.
The receiver unit is connected to the AC line
without a protective earth (class 2).
without a protective earth (class 2).
Degree of protection
against electric shock
against electric shock
Transducer - SF1-US : Class CF applied parts
Transducer - SF1-TOCO : Class CF applied
parts
Transducer - SF1-TOCO : Class CF applied
parts
Mode of operation.
Continuous
Degree of protection
against harmful ingress
of particles and/or water.
against harmful ingress
of particles and/or water.
Receiver - SF1-SL : IP30
Transducer - SF1-US : IPX8 (1m for 16Hours)
Transducer - SF1-TOCO : IPX8 (1m for 16Hours)
Transducer - SF1-US : IPX8 (1m for 16Hours)
Transducer - SF1-TOCO : IPX8 (1m for 16Hours)
Degree of safety of
application in the
presence of a fl ammable
anaesthetic
application in the
presence of a fl ammable
anaesthetic
Equipment not suitable for use in the presence of
a FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR, OXYGEN OR NITROUS OXIDE
a FLAMMABLE ANAESTHETIC MIXTURE WITH
AIR, OXYGEN OR NITROUS OXIDE
9.2 Receiver Unit (SF1-EUR / SF1-SL)
RF-Receiver
ISM : 434.05 - 434.7 MHz Band, 26 channels
WMTS : 608.0375 - 612.9875 MHz Band, 100 channels
WMTS : 608.0375 - 612.9875 MHz Band, 100 channels
Channel selection Rotary switch
Supply voltage
100V…240V, 50…60Hz, 10VA
Stand by power
consumption
consumption
<0.8W
Antenna
Gainfl ex
Transducer
Charging
Charging
Two docking and charging bays for SF1-US and SF1-
TOCO
TOCO
Charging time
Maximum: 3 hours, regulated
Operation
Automatic, no controls
Indicators
- Transmission US transducer
- Transmission TOCO transducer
- Battery capacity during operation
- Battery capacity during charging
- Battery charging established
- Transmission TOCO transducer
- Battery capacity during operation
- Battery capacity during charging
- Battery charging established
Size
250 x 200 x 85mm
Weight
1Kg