Analysis
pected which will point to the defective component. Rather 
exact results may be achieved if the measurements are 
compared to those of intact components. This is especially 
true for semiconductors. The polarity of diodes or transis-
tors can thus be identified if the lettering or marking is mis-
sing. Please note that with semiconductors changing the 
polarity will cause the display to rotate 180 degrees around 
the screen center. More important in practice is the quick 
determination of plain shorts and opens which are the most 
common causes of requiring service.
 
9 .5 .1 In-circuit tests
They are possible in many cases but deliver rarely clear 
results. By paralleling of real or complex impedances – es-
pecially if those are fairly low impedance at 50 Hz/200 Hz 
– there will be mostly great differences compared to 
individual components. If circuits of the same type have to 
be tested often (service), comparisons with intact circuits 
may help again. This is also quickly done because the intact 
circuit has not to be functional, also it should not be ener-
gized. Just probe the various test points with the cables of 
the component tester of the unit under test and the intact 
unit and compare the screen displays. Sometimes the unit 
under test may already contain an intact portion of the 
same type, this ist e.g. the case with stereo circuits, push-
pull circuits or symmetrical bridge circuits. In cases of doubt 
one side of the dubious component can be unsoldered.
9 .6  Digital Voltmeter
The digital voltmeter completely captures input data with 
the selected vertical sensitivity and the basic accuracy of 
the ADC. The digital voltmeter is independent from the 
capture settings and the post processing of the trans-
former values. Determine the values by means of the 
measurement interval to ensure the defined frequency
range of 20 Hz to 100 kHz. All measurements are perfor-
med by means of this measurement interval which ensu-
res that the DC portion can also be reliably captured for 
low-frequency signals.
The analysis function DIG. VOLTMETER can be activated 
by using the UTIL menu in the VERTICAL control panel. 
The digital voltmeter includes a 3-digit voltage display as 
Fig. 9.6: Digital Voltmeter.
7-segment display with the use of the analog channels. Four 
freely configurable measuring displays are possible where
two measurement value displays for channel 1 (CH1) and 
two measurement value displays for channel 2 (CH2) are 
available. For each channel, the measurement displays are 
divided into in a primary measurement location (soft menu 
key Channel 1 / Channel 2) and a secondary measurement 
location (soft menu key SECONDARY). The primary measure-
ment display will be shown in a large font, and the secondary 
measurement display in a small font. All measurement dis-
plays can be configured separately.
The following primary and secondary measurement displays 
are available:
 ❙
OFF: Measurement display deactivated
 ❙
DC: Mean
 ❙
DC RMS: RMS
 ❙
AC RMS: Standard Deviation
 ❙
CREST FACTOR: Crest factor (|X|max/XRMS)
 ❙
PEAK PEAK: Maximum-Minimum
 ❙
PEAK +: Maximum
 ❙
PEAK -: Minimum
Use the soft menu key POSITION to change the display 
position of the digital voltmeter on the screen. When the 
XY mode is activated, the display position of the digital volt-
meter is predefined and cannot be modified. Use the soft
menu key VOLTMETER OFF to deactivate the digital voltme-
ter function and to close the menu.