2.8.6 Logged Data Array Identification 
Each  sensor’s  logged  data  is  identified  by  an  array  ID  number.  For  the  primary  logged  data,  the  ID  is  the 
sensor  number  itself.  For  the  optional  supplementary  data  (min/max/deviation/flow),  the  array  ID  has  an 
offset added to the sensor number that it is derived from. These ID offsets are as follows: 
 
Minimum:  
+10 
Maximum: 
+20 
Deviation: 
+30 
Flow Rate 
+40 
 
For  example,  Sensor  4  has  been  configured  to  log  the  average  value,  plus  the  maximum  and  standard 
deviation. Three data arrays will be logged for this sensor at each logging interval with IDs of 4, 24 and 34 
respectively. In HydroTel™ 2000 these relate to point identifiers of 4/0, 24/0 and 34/0. 
 
 
2.8.7 Alarm Processing 
Each virtual sensor can be checked for two separate alarm conditions.  Each alarm has separate trigger and 
reset levels, as well as an activation delay or accumulation period depending upon the data type. The alarm 
on Sensor 1 can also be used to control the digital outputs.  See Sections 3.4.6 and 4.3.12 for further details 
on digital outputs. 
 
 
2.8.8 Real Time Clock & Calendar 
The iRIS has a non-volatile real time clock that can be set by the user either through a terminal (RS232 or 
UDP) or remotely via proprietary System DO
 commands from software such as HydroTel™ 2000 or iLink. 
To enable user adjustment to minimise clock drift a menu option is provided to set a compensation offset for 
fine control.  See Section 4.3.14.  
 
 
2.8.9 Security 
The  iRIS  320  can  be  configured  with  a  PIN  code  to  prevent  unauthorised  access  to  restricted  information 
through the LCD and keypad.  This is especially useful when the iRIS 320 is installed in a location where it is 
accessible to the general public.  
Note: This feature is not available on the iRIS 220 as it has no keypad/LCD interface.  
 
 
2.8.10 Gateway Communication  
The iRIS supports System DO
 gateway functionality between the GPRS/CDMA-1X network and the RS232 
serial  interface.    This  enables  the  unit  to  be  used  as  a  bridge  between  the  wide  area  GSM/CDMA-1X 
network and a localised network.  It is possible to connect a datalogger that does not have wireless capability 
such as the iQuest DS-4483 to the serial port of the iRIS and communicate with it via the gateway.  Also, by 
connecting a data radio to the unit’s serial port it is possible to communicate with several devices in a multi-
drop radio network from the GPRS/CDMA-1X network. 
 
The  gateway  transparently  redirects  System  DO
  packets  received  via  GPRS/CDMA-1X  from  address  0 
back out the serial port if they are not destined for the iRIS.  Conversely, packets received on the serial port 
from  any  address  and  directed  to  address  0  will  be  transparently  redirected  onto  the  GPRS/CDMA-1X 
network.  
 
To  enable  gateway  functionality  it  is  necessary  to  activate  the  RI  input  on  the  serial  port.    This  can 
easily  be  achieved  by  connecting  the  RI  and  DTR  pins  of  the  RS232  connector  together.    When  the  iRIS 
detects the presence of RI, it forces the serial port into System DO
 mode at 1200 baud. Refer to Appendix 
A – Radio Using the RS232 Interface for further information on using the RS232 port in this mode.