General Troubleshooting & Repair 
Teledyne API T703 Calibrator Operation Manual 
196  
 
11.2. USING THE ANALOG OUTPUT TEST CHANNEL 
The signals available for output over the T703’s analog output channel can also be used as diagnostic tools.  
See Section 6.9 for instruction on activating the analog output and selecting a function. 
Table 11-3:  Test Channel Outputs as Diagnostic Tools 
TEST 
CHANNEL 
DESCRIPTION ZERO 
FULL 
SCALE 
CAUSES  OF EXTREMELY  
HIGH / LOW READINGS 
NONE 
TEST CHANNEL IS TURNED OFF 
O3 PHOTO 
MEAS 
The raw output of the 
photometer during its 
measure cycle 
0 mV 
5000 mV* 
O3 PHOTO 
REF 
The raw output of the 
photometer during its 
reference cycle 
0 mV 
5000 mV 
If the value displayed is: 
-  >5000 mV: The UV source has become brighter.  Adjust the 
UV Detector Gain potentiometer. 
-  < 100mV – Bad UV lamp or UV lamp power supply. 
-  < 2000mV – Lamp output has dropped, adjust UV Preamp 
Board or replace lamp. 
 
If the value displayed is constantly changing: 
-  Bad UV lamp. 
-  Defective UV lamp power supply. 
- Failed I
2
C Bus. 
 
If the PHOTO REFERENCE value changes by more than 
10mV between zero and span gas: 
-  Defective/leaking M/R switching valve. 
O3 GEN 
REF 
The raw output of the 
O
3
 generator’s 
reference detector 
0 mV 
5000 mV 
Possible causes of faults are the same as OUTPUT FLOW from 
Table 11-2. 
OUTPUT 
FLOW 
Output flow rate 
(computed from 
regulator pressure). 
0 LPM 
6.000 LPM 
Possible causes of faults are the same as O3 GEN REFERENCE 
WARNING
 from Table 11-1 
SAMPLE 
PRESSURE 
The pressure of gas in 
the photometer 
absorption tube 
0 In-Hg-A 
40 In-Hg-A 
Check for Gas Flow problems.   
SAMPLE 
FLOW 
The gas flow rate 
through the photometer 
0 cm
3
/min 
1000 cc/m 
Check for Gas Flow problems.   
SAMPLE 
TEMP 
The temperature of gas 
in the photometer 
absorption tube 
C 70 
C 
Possible causes of faults are the same as PHOTO STEMP from 
Table 11-2 
PHOTO 
LAMP 
TEMP 
The temperature of the 
photometer UV lamp 
0 C
C 70 
C 
Possible failure of: 
-  Bench lamp heater  
-  Bench lamp temperature sensor 
-  Relay controlling the bench heater 
-  Entire Relay PCA 
-  I
2
C Bus 
- Hot” Lamp 
O3 LAMP 
TEMP 
The temperature of the 
O
3
 generator’s UV 
lamp 
0 mV 
5000 mV 
Same as PHOTO LAMP TEMP WARNING
 
CHASSIS 
TEMP 
The temperature inside 
the T703’s chassis 
(same as BOX TEMP
C 
Possible causes of faults are the same as BOX TEMP from Table 
11-2 
O3 PHOTO 
CONC 
The current 
concentration of O
3
 
being measured by the 
photometer. 
  - - - 
-  I
2
C Bus malfunction 
-  Gas flow problem through the photometer. 
-  Electronic failure of the photometer subsystems 
-  Failure or pressure / temperature sensors associated with the 
photometer 
-  Bad/incorrect Span Gas concentration 
-  Contamination of the Zero Air supply. 
-  Malfunction of the O
3
 generator. 
-  Internal A/D converter problem 