Teledyne MODEL T703 Manual Do Utilizador

Página de 287
General Troubleshooting & Repair 
Teledyne API T703 Calibrator Operation Manual 
198  
 
11.3.2.2. O
3
 Status LEDs 
 
D15 (Green) -  Photometer Lamp Heater
D16 (Yellow) – O
3
 Generator Lamp Heater
D2 (Yellow)  Dry (zero) Air Pump Status
D9 (Green) –  External Zero Air Valve Status
D1 (RED) 
Watchdog 
Indicator 
D7 (Green) Photometer Meas/Ref Valve Status
 
Figure 11-3: 
Relay PCA Status LEDS Used for Troubleshooting 
 
Table 11-5:  Relay PCA Status LED Failure Indications 
SIGNAL I/O PARAMETER 
LED FUNCTION 
ACTIVATED BY 
VIEW RESULT 
DIAGNOSTIC TECHNIQUE 
D2 
Yellow 
Status of AC 
powered Dry 
Air Pump 
ZERO_AIR_PUMP 
N/A 
Pump should start /stop 
 Failed pump 
  Failed AC Relay on Relay PCA 
  Failed Relay PCA 
  Faulty AC Power Supply (PS2) 
 Faulty Connectors/Wiring 
D7 
Green 
Photometer 
Meas/Ref 
Valve 
PHOTO_REF_VALVE 
N/A 
Valve should audibly change states.   
If not: 
 Failed Valve 
  Failed Relay Drive IC on Relay PCA 
  Failed Relay PCA 
  Faulty +12 VDC Supply (PS2) 
 Faulty Connectors/Wiring 
D9 
Green 
Status of DC 
powered 
Photometer 
Pump  
O3-PUMP-ON 
N/A 
Pump should start /stop 
 Failed pump 
  Failed Drive IC on Relay PCA 
  Failed Relay PCA 
  Faulty AC Power Supply (PS2) 
 Faulty Connectors/Wiring 
D15 
Green 
Photometer 
Heater Status 
PHOTO_LAMP_HEATER PHOTO_LAMP_TEMP 
D16 
Green 
O
3
 Generator 
Heater Status 
O3_GEN_HEATER O3_GEN_TEMP 
Voltage displayed should change.   
If not: 
 Failed Heater 
  Faulty Temperature Sensor 
  Failed AC Relay 
 Faulty Connectors/Wiring 
 
 
07223B DCN6378