11.3. USING THE INTERNAL ELECTRONIC STATUS LEDS 
Several LEDs are located inside the instrument to assist in determining if the calibrators CPU, I
2
C bus and Relay 
PCA are functioning properly. 
11.3.1. CPU STATUS INDICATOR 
DS5, a red LED, that is located on upper portion of the motherboard, just to the right of the CPU board, flashes 
when the CPU is running the main program loop.  After power-up, approximately 30 – 60 seconds, DS5 should 
flash on and off.  If characters are written to the front panel display but DS5 does not flash then the program files 
have become corrupted, contact customer service because it may be possible to recover operation of the 
calibrator.  If after 30 – 60 seconds neither DS5 is flashing and no characters have been written to the front 
panel display then the CPU is bad and must be replaced.  
 
Motherboard 
CPU Status LED 
 
Figure 11-2: 
CPU Status Indicator 
 
11.3.2. RELAY PCA STATUS LEDS 
There are seven LEDs located on the Relay PCA.  Some are not used on this model. 
11.3.2.1. I
2
C Bus Watchdog Status LEDs 
The most important is D1 (see, which indicates the health of the I
2
C bus.  
Table 11-4:  Relay PCA Watchdog LED Failure Indications 
LED 
Function 
Fault Status 
Indicated Failure(s) 
D1 
(Red) 
I
2
C bus Health 
(Watchdog Circuit) 
Continuously ON 
or 
Continuously OFF 
Failed/Halted CPU 
Faulty Motherboard or Relay PCA 
Faulty Connectors/Wiring between Motherboard or Relay 
PCA 
Failed/Faulty +5 VDC Power Supply (PS1) 
 
If D1 is blinking, then the other LEDs can be used in conjunction with DIAG Menu Signal I/O to identify hardware 
failures of the relays and switches on the Relay.  