Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Overview of Operating Modes 
 
 
83 
Table 4-2: Test Functions Defined 
DISPLAY PARAMETER  UNITS 
DESCRIPTION 
RANGE 
RANGE 
- - 
RANGE1 
RANGE2
 
PPB, PPM, 
UGM & MGM
The Full Scale limit at which the reporting range of the analyzer’s 
ANALOG OUTPUTS is currently set.   
THIS IS NOT the Physical Range of the instrument.  Refer to Section 
5.4 for more information. 
If DUAL or AUTO Range modes have been selected, two RANGE 
functions will appear, one for each range. 
STABIL 
STABILITY mV 
Standard deviation of SO
2
 Concentration readings.  Data points are 
recorded every ten seconds.  The calculation uses the last 25 data 
points. 
PRES 
SAMPLE 
PRESSURE 
in-Hg-A 
The current pressure of the sample gas as it enters the sample 
chamber, measured between the SO
and Auto-Zero valves. 
SAMP FL 
SAMPLE FLOW 
cm³/min 
(cc/m) 
The flow rate of the sample gas through the sample chamber. This 
value is not measured but calculated from the sample pressure. 
PMT 
PMT Signal 
mV 
The raw output voltage of the PMT. 
NORM PMT 
NORMALIZED 
PMT Signal 
mV 
The output voltage of the PMT after normalization for offset and 
temperature/pressure compensation (if activated). 
UV LAMP 
Source UV Lamp 
Intensity 
mV 
The output voltage of the UV reference detector. 
LAMP 
RATIO 
UV Source lamp 
ratio 
The current output of the UV reference detector divided by the reading 
stored in the CPU’s memory from the last time a UV Lamp calibration 
was performed. 
STR. LGT 
Stray Light 
ppb 
The offset due to stray light recorded by the CPU during the last zero-
point calibration performed. 
DRK PMT 
Dark PMT 
mV 
The PMT output reading recorded the last time the UV source lamp 
shutter was closed. 
DRK LMP 
Dark UV Source 
Lamp 
mV 
The UV reference detector output reading recorded the last time the 
UV source lamp shutter was closed. 
SLOPE 
SO
measurement  
Slope 
The sensitivity of the instrument as calculated during the last 
calibration activity.
 
The slope parameter is used to set the span 
calibration point of the analyzer. 
OFFSET 
SO
measurement 
Offset 
mV 
The overall offset of the instrument as calculated during the last 
calibration activity.  The offset parameter is used to set the zero point 
of the analyzer response. 
HVPS 
HVPS 
The PMT high voltage power supply. 
RCELL 
TEMP 
SAMPLE 
CHAMBER TEMP 
°C 
The current temperature of the sample chamber. 
BOX TEMP 
BOX 
TEMPERATURE 
°C 
The ambient temperature of the inside of the analyzer case. 
PMT TEMP 
PMT 
TEMPERATURE 
°C 
The current temperature of the PMT. 
IZS TEMP
1
 
IZS 
TEMPERATURE
1
 
°C 
The current temperature of the internal zero/span option. Only 
appears when IZS option is enabled. 
TEST
TEST SIGNAL
2
 
mV 
Signal of a user-defined test function on output channel A4.
 
TIME 
CLOCK TIME 
hh:mm:ss 
The current day time for DAS records and calibration events. 
1  
Only appears if Internal Gas Span Generator option is installed. 
2
 
Only appears if analog output A3 is actively reporting a test function.
 