Teledyne API - T100 UV Fluorescence SO2 Analyzer  
Overview of Operating Modes 
 
 
85 
4.1.2. WARNING MESSAGES 
The most common instrument failures will be reported as a warning on the analyzer’s 
front panel and through the COMM ports. Section 12.1.1 explains how to use these 
messages to troubleshoot problems. Section 12.1.3 shows how to view and clear 
warning messages. Table 4-3 lists all warning messages for the current version of 
software. 
Table 4-3: List of Warning Messages  
MESSAGE MEANING 
ANALOG CAL WARNING 
The instrument's A/D circuitry or one of its analog outputs is not calibrated. 
BOX TEMP WARNING 
The temperature inside the T100 chassis is outside the specified limits. 
CANNOT DYN SPAN 
Remote span calibration failed while the dynamic span feature was set to turned on 
CANNOT DYN ZERO 
Remote zero calibration failed while the dynamic zero feature was set to turned on 
CONFIG INITIALIZED 
Configuration was reset to factory defaults or was erased. 
DARK CAL WARNING 
Dark offset above limit specified indicating that too much stray light is present in the 
sample chamber. 
DATA INITIALIZED 
DAS data storage was erased. 
HVPS WARNING 
High voltage power supply for the PMT is outside of specified limits. 
IZS TEMP WARNING 
On units with IZS options installed: The permeation tube temperature is outside of 
specified limits. 
PMT DET WARNING 
PMT detector output outside of operational limits. 
PMT TEMP WARNING 
PMT temperature is outside of specified limits. 
RCELL TEMP WARNING 
Sample chamber temperature is outside of specified limits. 
REAR BOARD NOT DET 
The CPU is unable to communicate with the motherboard. 
RELAY BOARD WARN 
The firmware is unable to communicate with the relay board. 
SAMPLE FLOW WARN 
The flow rate of the sample gas is outside the specified limits. 
SAMPLE PRESS WARN 
Sample pressure outside of operational parameters. 
SYSTEM RESET 
The computer was rebooted. 
UV LAMP WARNING 
The UV lamp intensity measured by the reference detector reading too low or too 
high 
 
To view and clear warning messages, press: 
 
 SAMPLE           RANGE = 500.000 PPB         SO2 =XXX.X  
 
  TEST
             CAL 
 
      MSG
      
CLR
     
SETUP
 
Make sure warning messages are 
not due to real problems. 
Press CLR to clear the current 
message. 
If more than one warning is active, the 
next message will take its place 
Once the last warning has been 
cleared, the analyzer returns to 
SAMPLE
 mode. 
 SAMPLE 
       
RANGE=500.000 PPM
        SO2=XXX.X 
 
 < TST  TST >  CAL 
 
       
MSG
      
CLR
     
SETUP
 
SAMPLE 
         
 HVPS  WARNING
          
 SO2=XXX.X 
 
  TEST             CAL 
 
       MSG
      CLR     
SETUP
 
TEST 
ignores warning messages 
MSG 
activates warning 
messages.   
<TST  TST
> buttons replaced with 
TEST
 button 
NOTE: 
If the warning message persists 
after several attempts to clear it, 
the message may indicate a real 
problem and not an artifact of the 
warm-up period. 
 
Figure 4-3: Viewing and Clearing T100 WARNING Messages 
06807C DCN6650