Overview of Operating Modes 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
84 
To view the TEST Functions press the following button sequence: 
 
Figure 4-2: Viewing T100 TEST Functions 
I
MPORTANT
 
I
MPACT ON 
R
EADINGS OR 
D
ATA
 
A value of “XXXX” displayed for any of the TEST functions indicates an 
out-of-range reading or the analyzer’s inability to calculate it. All 
pressure measurements are represented in terms of absolute pressure. 
Absolute, atmospheric pressure is 29.92 in-Hg-A at sea level. It 
decreases about 1 in-Hg per 300 m gain in altitude. A variety of factors 
such as air conditioning and passing storms can cause changes in the 
absolute atmospheric pressure. 
06807C DCN6650