Setup Menu 
Teledyne API – Model T300/T300M CO Analyzer 
136 
 
5.9.3.11. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION 
To configure the analyzer’s external analog inputs option, define for each channel
  gain (number of units represented by 1 volt) 
 offset 
(volts) 
  engineering units to be represented in volts (each press of the touchscreen 
button scrolls the list of alphanumeric characters from A-Z and 0-9) 
  whether to display the channel in the Test functions 
To access and adjust settings for the external Analog Inputs option channels press: 
 
 
 
 
DIAG AIO
          
XIN1 GAIN:1.00V/V  
 
               SET>    EDIT 
 
  
 
EXIT 
 
 
 
DIAG
 
       
ANALOG I / O CONFIGURATION  
 
PREV 
NEXT 
      
          ENTR 
 
 EXIT 
 DIAG AIO
 
AOUTS CALIBRATED: NO
  
 
   < SET  SET>    CAL 
 
  
 
EXIT 
DIAG AIO
 
        
XIN1 GAIN:1.00V/V 
 
   +       0          0        1         .0          0         ENTR   EXIT 
Press to change 
Gain value 
 DIAG AIO
           
XIN1:1.00,0.00,V,OFF  
 
   < SET  SET>    EDIT 
 
  
 
EXIT 
Press SET> to scroll to the first 
channel. Continue pressing SET> 
to view each of 8 channels.  
Pressing ENTR records the new setting 
and returns to the previous menu. 
Pressing EXIT ignores the new setting and 
returns to the previous menu.
Press EDIT at any channel 
to to change Gain, Offset, 
Units and whether to display 
the channel in the Test 
functions (OFF/ON).  
 DIAG AIO
          
XIN1 OFFSET:0.00V  
 
   < SET
 
   SET>    EDIT 
 
  
 
EXIT 
 DIAG AIO
          
XIN1 UNITS:V  
 
   < SET    SET>    EDIT 
 
  
 
EXIT 
 DIAG AIO
          
XIN1 DISPLAY:OFF  
 
   < SET                 EDIT 
 
  
 
EXIT 
 
Figure 5-7.  DIAG – Analog Inputs (Option) Configuration Menu 
 