Enhancing Measurement Accuracy
A
Installation and User’s Guide
115
 
Performing a Coaxial System Measurement Calibration
Complete the following steps to perform a coaxial system measurement 
calibration and increase the accuracy of your device measurements. Modify 
these instructions as necessary to perform an in- fixture or on wafer 
measurement calibration.
To perform the system measurement calibration
Step
Action
Note
Preset the network 
analyzer.
On the network analyzer, press Preset.
Open the Calibration 
Wizard and choose a 
guided calibration.
In the network analyzer’s Calibration menu, 
select Calibration Wizard... .
In the Calibration Wizard dialog box, select 
Use Mechanical Stds... in the Guided 
Calibrations group box.
Choose the DUT 
connector types.
In the Guided Calibration: Select DUT 
Connectors dialog box, select the 
appropriate DUT connector type for the 
connections the Port 1 and Port 2 bias 
networks.
Click Next.
Choose the calibration kit. In the Guided Calibration: Select Version 2 
Cal Kits dialog box, select the appropriate 
calibration kit (for example 85056D) for the 
Port 1 and Port 2 bias network.
Click Next.