Installation and User’s Guide
A
Enhancing Measurement Accuracy
 
If you encounter a problem
Inspect the connectors on the load, open, short, and through standards 
and the connectors on the bias networks and the semi- rigid cables.
Clean the connectors if necessary. See step 2 in the section titled 
If connectors are damaged, replace the standard or cable.
Ensure that the standards meet their published specifications.
Calibrate the network 
analyzer through the bias 
networks, cables, and 
adapters using the 
standard open, short, 
load, and through.
Follow the prompts to connect the standard 
open to the Port 1 bias network RF/DC 
OUTPUT.
Click Measure.
Follow the prompts to connect the standard 
short to the Port 1 bias network RF/DC 
OUTPUT.
Click Measure.
Follow the prompts to connect the standard 
broadband load to the Port 1 bias network 
RF/DC OUTPUT.
f
Click Measure.
Repeat actions a through f for the Port 2 
calibration measurements.
Follow the prompts to connect the standard 
through between the Port 1 and Port 2 bias 
networks RF/DC OUTPUT. connectors.
i
Click Measure.
j
When the measurements are done, in the 
Standards Measured dialog box, click Next.
In the Guided Calibration Completed dialog 
box, choose the No, Finish Now radio 
button and click Finish.
Save the calibration and 
instrument state data to 
the C:/Program Files/
Agilent /Network 
Analyzer/Documents
 
folder.
In the File menu, select Save As... .
In the Save As dialog box, using the 
keyboard or by clicking Edit File Name
enter a file name (for instance, 
MY_CAL.CST
), then click OK.
This saves the calibration and 
instrument state data as a file (named 
for instance, MY_CAL.CST) in the 
network analyzer’s operating system 
C:/Program Files/Agilent/
Network Analyzer/Documents
 folder.
To perform the system measurement calibration (continued)
Step
Action