DC Subsystem Functional Verification Test
B
Installation and User’s Guide
121
 
Performing the DC Subsystem Functional Verification Test
Complete these steps to run the self- test and verify the operation the DC 
subsystem.
For Agilent 4156C subsystems:
If you encounter a problem
Refer to the Agilent 4156C precision semiconductor parameter analyzer 
manual titled, “If You Have a Problem” for an explanation of the error 
codes. See the chapter titled, “If Errors Occur When You Perform 
Self- Calibration or Diagnostics.”
Refer to the troubleshooting information in the service manual for the 
Agilent 4156C precision semiconductor parameter analyzer.
To perform the Agilent 4156C precision semiconductor parameter analyzer self-test
Step
Notes
Disconnect all cables from the 
measurement terminals on the 
4156C rear panel.
Connect the 4156C to line power.
Switch on the instrument line 
power switch.
Wait 1 hour before continuing to 
Press System > [CALIB/DIAG].
The analyzer displays the SYSTEM: 
SELF-CALIBRATION/DIAGNOSTICS 
screen.
Press [DIAG SELFTST ALL].
This begins the calibration and 
self-test process.
PASSFAIL, or DONE appear in the 
STATUS column.
If a failure occurs, an error code is 
displayed in the ERROR column. See 
If no errors occur, DIAG SELF-TEST 
ALL: PASS
 appears in the lower 
left-hand corner of the display.