DC Subsystem Functional Verification Test
B
Installation and User’s Guide
121
Performing the DC Subsystem Functional Verification Test
Complete these steps to run the self- test and verify the operation the DC
subsystem.
subsystem.
For Agilent 4156C subsystems:
If you encounter a problem
✔
Refer to the Agilent 4156C precision semiconductor parameter analyzer
manual titled, “If You Have a Problem” for an explanation of the error
codes. See the chapter titled, “If Errors Occur When You Perform
Self- Calibration or Diagnostics.”
manual titled, “If You Have a Problem” for an explanation of the error
codes. See the chapter titled, “If Errors Occur When You Perform
Self- Calibration or Diagnostics.”
✔
Refer to the troubleshooting information in the service manual for the
Agilent 4156C precision semiconductor parameter analyzer.
Agilent 4156C precision semiconductor parameter analyzer.
To perform the Agilent 4156C precision semiconductor parameter analyzer self-test
Step
Notes
1 Disconnect all cables from the
measurement terminals on the
4156C rear panel.
4156C rear panel.
2 Connect the 4156C to line power.
3 Switch on the instrument line
power switch.
4 Wait 1 hour before continuing to
5 Press System > [CALIB/DIAG].
•
The analyzer displays the SYSTEM:
SELF-CALIBRATION/DIAGNOSTICS
screen.
SELF-CALIBRATION/DIAGNOSTICS
screen.
6 Press [DIAG SELFTST ALL].
•
This begins the calibration and
self-test process.
self-test process.
•
PASS, FAIL, or DONE appear in the
STATUS column.
STATUS column.
•
If a failure occurs, an error code is
displayed in the ERROR column. See
displayed in the ERROR column. See
•
If no errors occur, DIAG SELF-TEST
ALL: PASS appears in the lower
left-hand corner of the display.
ALL: PASS appears in the lower
left-hand corner of the display.