B
DC Subsystem Functional Verification Test
 
For Agilent E5260A/70B DC subsystems:
If you encounter a problem
Refer to the Agilent E5260A/70B precision parametric measurement 
solution user’s guide for an explanation of the error codes. See the 
section titled “Error Codes” in the chapter titled, “If You Have a 
Problem.”
Refer to the troubleshooting information in the Agilent E5260A/70B 
precision parametric measurement solution service manual.
To perform the Agilent E5260A/70B precision parametric measurement solution self-test
Step
Notes
Disconnect all cables from the 
measurement terminals on the 
E5260A/70B front panel.
Connect the E5260A/70B to line 
power.
Switch on the instrument line 
power switch.
Wait 1 hour before continuing to 
Press Shift > Menu.
Move the cursor to SELFTEST and 
press Enter. Move the cursor to 
EXECUTE and press Enter.
Use the arrow keys to select ALL 
and press Enter. At the completion 
of the test, press Exit three times to 
exit the setup menu.
To display the results, move the 
cursor to SELFTEST and press 
Enter. Move the cursor to RESULT 
and press Enter.
Use the arrow keys to select 
FRAME or Slot n.
The test result is displayed for each 
item.