Installation and User’s Guide
2
Installing the System
 
Agilent 4156C Systems with Agilent 41501B Expander Box
To connect the bias networks
Step
Action
Notes
For Agilent 4156C with 
41501B expander box 
systems, refer to the 
following figure.
Connect the triaxial 
cables from the 4156C to 
the bias networks.
Connect the triaxial cables from the 4156C 
HRSMU1 FORCE and SENSE connectors to 
the DC FORCE and DC SENSE connectors 
on the port 1 bias network.
Connect the triaxial cables from the 41501B 
HPSMU FORCE and SENSE connectors to 
the DC FORCE and DC SENSE connectors 
on the port 2 bias network.
From the rear of the system, route the 
cable through the feedthrough panel.
The port 1 bias network is the 11612V 
K11.
The port 2 bias network is the 11612V 
K21.
Refer to 
Connect and route the 
triaxial cable from 4156C 
HRSMU3 FORCE 
connector to the port 2 
bias network.
From the rear of the system, connect one 
end of the triaxial cable to the GNDU 
connector on the 41501B rear panel.
From the rear of the system, route the cable 
through the upper feedthrough panel.
Connect the triaxial cable to the port 2 bias 
network GNDU connector.
The triaxial cable model number is 
16494A Option 002.
Leave the GNDU connector on the port 
1 bias network open.
Refer to