Installation and User’s Guide
2
Installing the System
 
Agilent E5260A/70B Systems
To connect the bias networks
Step
Action
Notes
For Agilent E5260A/70B 
systems, refer to the 
following figure.
Connect the triaxial 
cables to the E5260A/70B 
SMUs.
*
From the front of the system rack, connect 
triaxial cables to the front panel high power 
SMU FORCE and SENSE outputs on the 
E5260A/70B.
Connect a triaxial cable to the front panel 
medium power SMU FORCE and SENSE 
outputs on the E5260A/70B.
From the front of the system, route the 
cable through the upper feedthrough panel.
From the rear of the system, route the 
HPSMU cable through the port 2 hole in the 
lower feedthrough panel. 
From the rear of the system, route the 
MPSMU cable through the port 1 hole in 
the lower feedthrough panel.
The triaxial cable model number is 
16494A Option 002.
The high speed high power SMU is the 
Agilent E5290A.
The high speed medium power SMU is 
the Agilent E5291A.
The high power SMU is the Agilent 
E5280A.
The medium power SMU is the Agilent 
E5281A.