FIGURE 9-3 Recalibration Screen (Part I)
4. Touch the red targets on the screen as instructed. Three calibration targets will be 
displayed.
NOTE:
The messages in the Recalibrate Touch Screen display are 
always displayed in the English language.
NOTE:
During the recalibration process, the patient tiles, menu bar, 
and sidebar buttons will be replaced by the recalibration 
screen. 
5. Once the three targets have been touched, the message Touch the screen. Does the 
cursor follow your finger? is displayed (shown in FIGURE 9-5). 
Select either the 
 or 
 button.
• Select the 
 button if the targets moved as they were selected. Once selected, the 
screen is recalibrated.
• Select the 
 button if the targets did not move as they were selected. Once selected, 
the recalibration process begins again.
NOTE:
When the central station, ViewStation, or WorkStation uses 
a dual display configuration, recalibration will affect both 
touch screen displays. The first touch screen display will be 