9.3 
Recharging the battery 
Battery power display 
On the front of the IntegraMouse Plus there is a display consisting of two light-emitting diodes 
(LEDs) in the form of a battery, which displays the battery power and the charging progress. 
LED 1 
LED 2 
Battery power  Note 
green 
green 
100% - 71% 
Battery fully charged 
green 
Off 
70% - 21% 
Battery OK, charging not yet required 
flashing orange 
Off 
below 20% 
Low battery, please charge 
 
Table 3: Battery power display 
 
Charging 
The charging of the battery occurs via the power pack included or by connection to the 
computer via the USB cable. In both cases, use the mini USB socket on the reverse of the 
IntegraMouse Plus. Plugging in is only possible in one direction. The charging process 
begins immediately after plugging in and takes 2-3 hours. Progress is indicated on the battery 
display: 
LED 1 
LED 2 
Battery power  Note 
orange 
Off 
Battery is 
being charged 
Leave power pack plugged in until charging 
process is completed. 
green 
green 
Battery fully 
charged 
Power pack can be unplugged 
 
Table 4: Charging process display 
 
 
Only use the power pack or the USB cable supplied to charge or operate the 
mouse. Even when you have another power pack with mini USB available (e.g. 
from a mobile telephone), do not use this to charge or operate the IntegraMouse 
Plus.  
 
 
Never change the battery yourself! If you suspect a problem with the battery, 
contact the manufacturer or your dealer. 
9.4 
Metrological testing 
This product - especially when being used in a clinical environment - should regularly be 
checked for safety. Regular safety testing - at the most every 24 months - can be carried out 
by the operator, by authorized service personnel or by the manufacturer. Responsibility for 
compliance lies with the operator of the IntegraMouse Plus.