Instructions for Use | Therapy Modes and Controls 
 
 
 
Control Modes 
20 
 
 
A/C-VC: Assisted/Control-Volume Control 
D
ESCRIPTION
:
 
 
The A/C-VC mode provides volume-controlled mandatory and assist breaths. When the Trigger Type is set to Off, both triggering and cycling are 
performed by the ventilator. When the Trigger Type is not Off, then the trigger can be performed by the vent or patient and the cycle is always 
performed by the ventilator. To deliver the set volume in the set time, the ventilator alters the flow rate. The flow pattern setting defines the shape 
of the flow delivery pattern. 
S
ETTINGS
:
 
 
Setting Name 
Description 
Tidal Volume 
Set inspiratory volume 
PEEP 
Positive end expiratory pressure 
Inspiratory Time 
Length of the inspiratory phase 
Breath Rate 
Minimum rate of mandatory breaths per minute 
Flow Pattern 
Sets the shape of the waveform as a ramp or square 
Trigger Type 
• 
Auto-Trak 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
• 
Off 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the 
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger. 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the 
device cycles to expiration. 
 
S
ETTABLE ALARMS
• 
Circuit disconnected 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
• 
High inspiratory pressure 
• 
Low inspiratory pressure 
I
LLUSTRATION