Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Therapy Modes and Controls 
 
 
 
Spontaneous Modes 
22 
 
 
PSV: Pressure Support Ventilation 
D
ESCRIPTION
:
 
 
PSV mode is patient-triggered, pressure-limited, and flow-cycled. With this strategy, breaths are assisted by a set inspiratory pressure that is 
delivered until inspiratory flow drops below a set threshold.  
In the PSV mode, the ventilator delivers spontaneous, pressure-supported, breaths and user-initiated mandatory breaths. The ventilator functions 
as a demand flow system, with the patient triggering breaths and determining their timing and volume. The ventilator can support the breaths with 
the set pressure support.  
The Pressure Control setting defines the applied pressure above PEEP. The patient determines the breath timing. As in other dual limb modes, you 
also set PEEP, inspiratory trigger, and O
2
. It is recommended that you set backup ventilation in PSV mode. 
S
ETTINGS
Setting Name 
Description 
Pressure Support  Target pressure that the device delivers during the inspiratory phase of a spontaneous breath 
PEEP 
Positive end expiratory pressure 
Rise Time 
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory 
pressure setting when the breath is triggered. 
Trigger Type 
• 
Auto-Trak 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the 
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the 
device cycles to expiration. 
 
S
ETTABLE ALARMS
:
 
 
• 
Circuit disconnected 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
I
LLUSTRATION
: