Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Therapy Modes and Controls 
 
 
 
Spontaneous Modes 
21 
 
Spontaneous Modes 
CPAP: Continuous Positive Airway Pressure 
D
ESCRIPTION
:
 
 
In CPAP mode, the pressure delivered to the patient during both inhalation and exhalation is the CPAP pressure setting. All breaths in this mode are 
spontaneous breaths. The ventilator monitors inspiratory and expiratory tidal volume. 
S
ETTINGS
:
 
 
Setting Name 
Description 
CPAP 
Continuous positive airway pressure range 
Trigger Type 
• 
Auto-Trak 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the 
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity 
setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger. 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set 
point, the device cycles to expiration. 
 
S
ETTABLE ALARMS
:
 
 
• 
Circuit disconnected 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
I
LLUSTRATION