Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Therapy Modes and Controls 
 
 
 
Mixed Modes 
24 
 
 
SIMV-PC: Synchronous Intermittent Mandatory Ventilation- Pressure Control 
D
ESCRIPTION
:
 
 
SIMV-PC mode is a pressure control mode that provides a mixture of mandatory and spontaneous breaths. SIMV-PC mode guarantees one 
mandatory breath in each cycle. Spontaneous breaths can be delivered with pressure support. The breath rate determines the length of the cycle. 
The first phase of the cycle is reserved for synchronizing a mandatory breath with patient effort. If the patient triggers a breath during this phase of 
the cycle, the ventilator delivers a synchronized mandatory breath. If a patient does not trigger a breath during the mandatory phase of the cycle, 
then the ventilator delivers a mandatory breath. Breaths triggered by the patient after the mandatory breath in the cycle are spontaneous breaths. 
This process is repeated at the start of every cycle.  
S
ETTINGS
:
 
 
Setting Name 
Description 
Pressure Control 
Defines the applied pressure for all breaths  
Pressure Support  Target pressure that the device delivers during the inspiratory phase of a spontaneous breath 
PEEP 
 
Positive end expiratory pressure 
Positive pressure maintained in the patient circuit during exhalation: must be less than or equal to the 
pressure setting. 
Inspiratory Time 
For a mandatory breath, length of the inspiratory phase 
Rise Time 
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory pressure 
setting when the breath is triggered. 
Breath Rate 
Minimum rate of mandatory breaths per minute 
Trigger Type 
• 
Auto-Trak 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the patient’s 
inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the 
device cycles to expiration. 
S
ETTABLE ALARMS
• 
Circuit disconnect 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
I
LLUSTRATION