Instructions for Use | Therapy Modes and Controls 
 
 
 
Therapy Features 
25 
 
 
SIMV-VC: Synchronous Intermittent Mandatory Ventilation- Volume Control 
D
ESCRIPTION
:
 
 
Similar to SIMV-PC, but with volume control. 
S
ETTINGS
:
 
 
Setting Name 
Description 
Tidal Volume 
Target gas volume that the device delivers during a spontaneous breath 
Pressure Support  Target pressure that the device delivers during the inspiratory phase of a spontaneous breath 
PEEP 
 
Positive end expiratory pressure 
Positive pressure maintained in the patient circuit during exhalation: must be less than or equal to the 
pressure setting. 
Inspiratory Time 
For a mandatory breath, length of the inspiratory phase 
Rise Time 
Time required for the ventilator to change from the expiratory pressure setting to the inspiratory 
pressure setting when the breath is triggered. 
Breath Rate 
Minimum rate of mandatory breaths per minute 
Flow Pattern 
Sets the flow-pressure waveform 
Trigger Type 
• 
Auto-Trak 
• 
Sensitive Auto-Trak (passive circuits only) 
• 
Flow Trigger (Passive, Active PAP, Active Flow, or Dual Limb circuits) 
Flow Trigger 
Sensitivity 
This control is available when the trigger type is Flow Trigger. The flow trigger initiates when the 
patient’s inspiratory effort creates a flow equal to or greater than the flow trigger sensitivity setting. 
Flow Cycle 
Sensitivity 
This control is available when the trigger type is Flow Trigger 
As flow begins to decrease during inspiration, if the patient flow is less than the flow cycle set point, the 
device cycles to expiration. 
 
S
ETTABLE ALARMS
• 
Circuit disconnect 
• 
High tidal volume 
• 
Low tidal volume 
• 
High minute ventilation 
• 
Low minute ventilation 
• 
High respiratory rate 
• 
Low respiratory rate 
• 
High inspiratory pressure 
• 
Low inspiratory pressure 
I
LLUSTRATION
 
See SIMV-PC. 
Therapy Features 
The following features are available in addition to the therapy modes.  
Backup Ventilation Enable 
D
ESCRIPTION
 
Set the device to deliver ventilator-initiated breaths when patient-initiated breaths are not detected, based on the Apnea alarm interval. When you 
turn Backup Ventilation on, set an Apnea interval in the alarm settings tab. Within the apnea interval; if no breaths are triggered by the patient, the 
ventilator delivers breaths at the set pressure or volume based on the Backup Rate. When an Apnea alarm occurs, the ventilator automatically 
starts backup ventilation.  When two consecutive patient-initiated breaths are detected, the ventilator automatically reverts to patient-initiated 
breaths.  
Backup ventilation settings take precedence over standard therapy mode settings.