Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Device Options 
 
 
 
Calibration 
67 
 
 
Leak Test 
A leak test is available for times when you want to perform only a leak test and not complete an entire circuit calibration. 
Prerequisites: 
• 
Ensure the prescription is for an active circuit: 
Active Flow 
Active PAP  
Dual Limb 
• 
Remove the patient interface from the circuit. 
• 
Block the end of the circuit where the patient interface would be. 
• 
If the circuit includes a heated humidifier, ensure the humidifier chamber is filled.  
• 
Ensure the external active exhalation valve is properly assembled and connected. 
 
To perform a leak test: 
1.  In the menu bar, tap the Options icon. 
 
In the Options window, tap Calibration and Setup
2.  In the Calibration & Setup window, tap Leak Test
3.  Review the prerequisites and then tap Start
4.  The system performs the test. The results appear in the test progress pane. If the test: 
• 
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap the cancel button at the 
top of the window. 
• 
Passes: tap OK
 
O
2
 Sensor Calibration 
Prerequisites: 
• 
Ensure the patient circuit is connected. 
• 
Remove the patient interface from the circuit. 
• 
Remove any passive exhalation device. 
• 
Connect high-pressure O
2
.  
• 
Ensure that low-flow O
2
 is not connected. 
 
To calibrate the O
2
 sensor: 
1.  In the menu bar, tap the Options icon. 
 
2.  In the Options window, tap Calibration and Setup
3.  In the Calibration & Setup window, tap O
2
 Sensor Calibration
4.  Review the prerequisites and then tap Start
5.  The system performs the Circuit Flush test. The results appear in the progress pane. If the test:  
• 
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit.  
• 
Passes: go to the next step. 
6.  Block the end of the circuit and then tap Continue
7.  The system performs the 21% Calibration test. The results appear in the progress pane. If the test:  
• 
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit
• 
Passes: go to the next step. 
8.  If you are using an oxygen-blending module, confirm that high-pressure O
2
 is connected and the circuit outlet is blocked 
and then tap Continue. Otherwise, the test is complete.