Respironics Inc. 1127941 Manual Do Utilizador

Página de 100
 
 
Instructions for Use | Device Options 
 
 
 
Data Transfer 
69 
 
4.  In the Data Transfer window, tap the procedure you would like to perform and then follow the prompts on the screen:  
 
Function 
Description 
Export Data - USB 
 
Export details about the patient’s therapy to a storage device connected to the USB 
port. Data is identified by the device serial number. 
Data type 
Data stored 
Changes to the ventilator, alarms, prescriptions, accessories 
and oxygen 
6 months 
30-second averages of each therapy parameter related to 
the therapy provided to the patient 
Waveform data including pressure, flow, volume, and total 
leak 
31 days 
Patient monitors: 
CO
2
 monitor data 
Oximeter data 
FiO
2
 monitor data 
 
Export Data - Bluetooth 
Export details about the patient’s therapy to a storage device connected through 
Bluetooth. Data is identified by the device serial number. 
Data type 
Data stored 
Changes to the ventilator, alarms, prescriptions, accessories 
and oxygen 
6 months 
30-second averages of each therapy parameter related to 
the therapy provided to the patient 
Waveform data including pressure, flow, volume, and total 
leak 
31 days 
Patient monitors: 
CO
2
 monitor data 
Oximeter data 
FiO
2
 monitor data 
 
Export Alarm & Event Log 
Export the alarm and event logs. If two storage devices are connected, the logs are 
saved to the device that was connected first.  
Install Prescriptions 
Installs prescriptions that were previously exported.  
Install Software Update 
 
Ensure the version number that you are installing is higher than the current software 
version. After installing the update, the system automatically restarts. 
 
As data is transferred, a data transfer icon appears in the Status Bar.  
USB Data Transfer Icon     
                       Bluetooth Data Transfer icon