Instructions for Use | Device Options 
 
 
 
Data Transfer 
68 
 
9.  The system performs the test. The results appear in the test progress pane. If the test: 
• 
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap Quit. To deactivate the FiO
2
 
sensor, turn off the FiO
2
 sensor in Device Options, Device Settings. 
• 
Passes: tap OK
 
CO
2
 Sensor Adapter Zero  
Perform this task when installing a CO
2
 sensor (capnograph)  
To establish a baseline CO
2
 level:  
1.  Place the CO
2
 sensor onto a clean and dry airway adapter that is exposed to room air, but is away from all CO
2
 sources 
including the ventilator, your breath, and the patient’s breath. 
2.  In the menu bar, tap the Options icon.  
3.  In the Options window, tap Calibration and Setup. 
4.  In the Calibration & Setup window, tap CO
2
 Sensor Adapter Zero
5.  Review the prerequisites and then tap Start
6.  The system performs the test. The results appear in the test progress pane. If the test: 
• 
Fails: Review the failure reasons. If you want to try again, tap Retest; otherwise, tap the Quit
• 
Passes: tap OK
 
Data Transfer 
Use data transfer functions to import and export various sets of data. When working with patient data, ensure the data is kept secure. When 
working with an external storage device that contains patient information, such as a USB flash drive, ensure you delete that data by reformatting 
the storage device. 
To access the data transfer functions: 
1.  Connect a data storage device to the USB port on the Service Panel. 
2.  In the menu bar, tap the Options icon. 
 
3.  In the Options window, tap Data Transfer