Instructions for Use | Technical Specifications 
 
 
 
EMC Information 
95 
 
Immunity Test 
IEC 60601 Test Level 
Compliance Level 
Electromagnetic Environment - Guidance 
IEC 61000-4-8 
Electromagnetic 
compatibility 
(EMC). Part 4-8: 
Testing and 
measurement 
techniques. Power 
frequency (50/60 
Hz) magnetic field 
immunity test 
30 A/m  
30 A/m  
Power frequency magnetic fields should be at levels 
characteristic of a typical location in a typical hospital 
or home environment.  
NOTE: UT is the AC mains voltage prior to application of the test level.  
 
 
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity 
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it 
is used in such an environment. 
Immunity Test 
IEC 60601 Test Level Compliance Level 
Electromagnetic Environment - Guidance 
Conducted RF IEC 61000-
4-6 
Electromagnetic 
compatibility (EMC). Part 
4-6: Testing and 
measurement techniques. 
Immunity to conducted 
disturbances, induced by 
radio-frequency fields. 
3 Vrms 
150 kHz to 80 MHz 
3 Vrms 
150 kHz to 80 MHz 
Portable and mobile RF communications 
equipment should be used no closer to any part of 
the device, including cables, than the 
recommended 30 cm separation distance.  
6 Vrms 
Amateur Radio & 
ISM Bands between 
150 kHz and 80 MHz 
6 Vrms 
Amateur Radio & ISM 
Bands between 150 kHz 
and 80 MHz 
Radiated RF 
IEC 61000-4-3 
Electromagnetic 
compatibility (EMC) - Part 
4-3: Testing and 
measurement techniques 
- Radiated, radio-
frequency, 
electromagnetic field 
immunity test 
10 V/m 
80 MHz to 2.7 GHz 
10 V/m 
Up to 28V/m in 
telecommunication 
bands as specified in 
clause 8.10 of IEC 
60601-1-2 
Up to 28V/m in 
telecommunication bands 
as specified in clause 8.10 
of IEC 60601-1-2