93 
 
 
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions  
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it 
is used in such an environment. 
Emissions Test 
Compliance 
Electromagnetic Environment - Guidance 
RF emissions  
CISPR 11 Industrial, scientific and medical 
equipment. Radio-frequency disturbance 
characteristics. Limits and methods of 
measurement 
Group 1  
The device uses RF energy only for its internal function. 
Therefore, its RF emissions are very low and are not 
likely to cause any interference in nearby electronic 
equipment.  
RF emissions  
CISPR 11  
Class B  
The device is suitable for use in all establishments, 
including domestic establishments and those directly 
connected to the public low-voltage power supply 
network.  
Harmonic emissions  
IEC 61000-3-2 Electromagnetic compatibility 
(EMC). Part 3-2: Limits. Limits for harmonic 
current emissions (equipment input current 
smaller than or equal to 16 A per phase) 
Class A  
Voltage fluctuations/Flicker emissions  
IEC 61000-3-3 Electromagnetic compatibility 
(EMC) - Part 3-3: Limits - Limitation of voltage 
changes, voltage fluctuations and flicker in 
public low-voltage supply systems, for 
equipment with rated current ≤ 16 A per 
phase and not subject to conditional 
connection 
Complies  
Emission of Radio Frequency Energy  
RTCA/DO-160G Section 21  
Category M  
This device is suitable for use onboard commercial 
airplanes inside passenger cabin.