Guidance and Manufacturer’s Declaration - Electromagnetic Immunity  
This device is intended for use in the electromagnetic environment specified below. The user of this device should make sure it 
is used in such an environment. 
Immunity Test 
IEC 60601 Test Level 
Compliance Level 
Electromagnetic Environment - Guidance 
IEC 61000-4-2 
Electromagnetic 
compatibility 
(EMC). Part 4-2: 
Testing and 
measurement 
techniques. 
Electrostatic 
discharge 
immunity test 
±8 kV contact  
±2 kV, ±4 kV, ±8 kV, and 
 ±15 kV air  
±8 kV contact  
±2 kV, ±4 kV, ±8 kV, 
and 
 ±15 kV air 
Floors should be wood, concrete, or ceramic tile. If 
floors are covered with synthetic material, the 
relative humidity should be at least 30%.  
IEC 61000-4-4 
Electromagnetic 
compatibility 
(EMC). Part 4-4: 
Testing and 
measurement 
techniques. 
Electrical fast 
transient/burst 
immunity test 
±2 kV for power supply lines  
±1 kV for input-output lines  
±2 kV for supply 
mains  
±1 kV for 
input/output lines  
Mains power quality should be that of a typical home 
or hospital environment.  
IEC 61000-4-5 
Electromagnetic 
compatibility 
(EMC) - Part 4-5 
Testing and 
measurement 
techniques. Surge 
immunity test 
±1 kV line to ground  
±2 kV line to ground  
±1 kV line to line  
N/A  - this Class II 
device does not 
connect to earth 
ground 
Mains power quality should be that of a typical home 
or hospital environment.  
IEC 61000-4-11 
Electromagnetic 
compatibility 
(EMC). Part 4-11: 
Testing and 
measurement 
techniques. 
Voltage dips, short 
interruptions and 
voltage variations 
immunity tests 
O% U
0.5 cycle at 45 degree 
increments  
0% U
 T
 1 cycle 
70% U
 T
 25 cycles  
(30 cycles if US) 
0% U
 T
 5 sec  
O% U
 T
 0.5 cycle at 45 
degree increments  
0% U
 T
 1 cycle 
70% U
 T
 25 cycles  
(30 cycles if US) 
0% U
 T
 5 sec  
Mains power quality should be that of a typical home 
or hospital environment. If the user of the device 
requires continued operation during power mains 
interruptions, it is recommended that the device be 
powered from an uninterruptible power supply or a 
battery.