HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 4
101
CDMA Measurements - Screens and Control Fields
Gen Dir
Gen Mode
Gen Dir
This field sets the link direction of the CDMA signal. 
Fwd
 (forward) sets the CDMA generator to produce QPSK 
modulation to imitate a base station. The 
EQ Filter
 field is 
automatically set to 
EQ In
 to apply an equalizing filter to the I and 
Q paths as part of the baseband filtering, but it can be changed to 
Out
Rev
 (reverse) sets the CDMA generator to produce OQPSK 
modulation to imitate a mobile station. The 
EQ Filter
 field is 
automatically set to 
Out
 (and cannot be changed) to remove the 
equalizing filter from the I and Q paths. 
Screens on which this field is present
CDMA GENERATOR
Gen Mode 
This field selects the type of modulation information sent to the CDMA 
generator’s IQ modulator. 
Data
 allows data to be transmitted to simulate a traffic channel with 
no additive noise. (See the
.) 
Eb/No
 causes a combination of data and White Gaussian Noise 
(AWGN) noise to be transmitted. The value entered is the ratio 
between the energy of each information bit (E
b
) and the noise 
spectral density (N
o
), expressed in dB. 
Noise
 causes a signal modulated by White Gaussian Noise to be 
generated. The noise is band limited to 2 MHz and white inside a 
1.25 MHz bandwidth centered around the 
RF Gen Freq
. This 
selection removes the 
Data Source
Start Frame
, and 
# of 
Frames
 fields from the screen; those fields are only used with the 
Data
 and 
Eb/No
 entries. 
Screens on which this field is present
CDMA GENERATOR
Main Menu