HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 5
125
Analog Measurements - Screens and Control Fields
AFGen2 Freq
AFGen2 To
AFGen2 Freq 
This field sets the frequency for the second audio frequency sinewave 
generator.
Screens on which this field is displayed
RF GENERATOR
AFGen2 To
This field has two subfields: 
• The upper field sets the destination port for audio frequency 
generator 2 
FM
 -RF generator FM modulator 
AM
 -RF generator AM modulator
Audio Out
 -AUDIO OUT connector
• The lower field 
• sets FM modulation deviation if the upper field set to 
FM
 
• sets AM modulation depth if the upper field set to 
AM
• sets the amplitude of audio signal (volts RMS) at the AUDIO 
OUT connector if upper field is set to 
Audio Out
• turns off the AF generator if the 
Yes On/Off
 key is pressed.
Operating Considerations
Simultaneous FM and AM, using both AF generators is not allowed.
AF generators 1 and 2 are automatically turned off any time the CDMA 
screens are displayed. This prevents accidental amplitude or frequency 
modulation of the CDMA carrier. The AF generators are turned back on 
when you exit the CDMA screens (if they were previously turned on).
Screens on which this field is displayed
RF GENERATOR
Main Menu