Chapter 5 
Analog Measurements - Screens and Control Fields
Amplitude
Atten Hold
Amplitude
This field adjusts the amplitude of the RF generator. 
See for Atten Hold (below) for more information.
Screens on which this field is displayed
RF GENERATOR
SPEC ANL: RF generator controls menu
Atten Hold 
Attenuator hold prevents the fixed RF output attenuators from 
switching in and out, eliminating the loss of the output signal as the 
level is changed. This function is helpful when making squelch 
measurements. 
Operating Considerations
When this function is set to 
On
, the RF output level is restricted to a 
range around the present 
Amplitude
 setting. This range varies with 
the amplitude setting. Attempting to set an amplitude outside the 
allowed range results in an error message and beep (if the beeper is on). 
RF output level accuracy is greatly degraded outside the allowed range. 
Screens on which this field is displayed
RF GENERATOR
Main Menu