HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 5
127
Analog Measurements - Screens and Control Fields
Audio In Lo
Auto/Norm
Audio In Lo 
This field sets the AUDIO IN LO connector’s state. 
Gnd
 causes the center pin of the connector to be connected directly to 
chassis ground.
Float
 isolates the center pin of the connector from ground, 
providing a floating input to the AF analyzer.
600 To Hi
 establishes a 600-ohm impedance between the center 
pins of the AUDIO IN LO and AUDIO IN HI connectors. Also, the 
Ext Load R
 field is removed, since the load is now fixed to 600 
ohms.
Screens on which this field is displayed
AF ANALYZER
Audio Out 
Audio output coupling selects ac or dc coupling of the AF generator to 
the AUDIO OUT connector. 
Screens on which this field is displayed
RF GENERATOR
Auto/Norm 
This field specifies how the trigger level is set. 
Auto
 automatically triggers if a triggering signal is not detected 
within approximately 50 ms of the last trigger. 
Norm
 requires a specific triggering signal before triggering. 
Operating Considerations
Automatic triggering should be used for signals >20 Hz. Normal 
triggering should be used for signals 
20 Hz. 
Also, when measuring 
1 Hz signals, you should set the 
Scope To
 field 
in the AF ANALYZER screen to 
Input
 to provide dc coupling to the 
oscilloscope’s input. 
Screens on which this field is displayed
SCOPE: trigger controls menu
Main Menu