HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 5
133
Analog Measurements - Screens and Control Fields
FM Coupling
Freq (marker)
FM Coupling
This field alters the FM modulator to allow DCFM from internal and 
external modulation sources. This field also selects ac or dc coupling 
between the RF generator’s frequency modulator and the ANALOG 
MODULATION IN connector.
Operating Considerations
This field should be set to 
DC
 whenever the 
AFGen2 To
 field is set to 
FM
this provides better modulation response at low data rates. 
Screens on which this field is displayed
RF GENERATOR
Freq (marker)
Marker frequency displays the frequency at the marker’s present 
position on the spectrum analyzer. 
Screens on which this field is displayed
SPEC ANL: all control menus
Figure 5-1  Freq and Lvl Display Fields
Main Menu