HP Cell Phone Accessories E6380-90019 Manual Do Utilizador

Página de 228
Chapter 5
135
Analog Measurements - Screens and Control Fields
IF Filter
Input Atten
IF Filter 
This field selects the desired IF filter bandwidth for modulated signals 
being analyzed. 
Screens on which this field is displayed
RF ANALYZER
Input Atten
Input attenuation sets the amount of input attenuation for the RF 
IN/OUT and ANT IN connectors. This function controls two settings: 
• The upper field determines if you want the instrument to set the 
attenuation automatically (
Auto
), or if you want to set the value 
manually (
Hold
).    
• The lower field displays the present attenuation value, and is used to 
set the desired attenuation level when the upper area is set to 
Hold
Operating Considerations
Autoranging Interference.  After a signal is input, the RF autoranging 
function (
Auto
) takes a small amount of time to determine the required 
input attenuator setting. If your transmitter begins sending 
information the instant it transmits, the initial part of the demodulated 
signal may not appear on the oscilloscope. When trying to capture the 
initial modulation waveform of a signal on the oscilloscope, set the 
upper field to 
Hold
 and set the lower field to an appropriate level for 
the signal being decoded (start with 40 dB).
Setting the upper field to 
Hold
 also prevents the RF autoranging 
process from interrupting spectrum analyzer operation when a signal is 
first measured. This can be helpful when you need to see the signal the 
instant the source is input, but requires you to set the needed amount of 
input attenuation. 
Screens on which this field is displayed
RF ANALYZER
SPEC ANL: auxiliary controls menu
Main Menu