Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
122 
To enter the signal I/O test mode, press:  
 
SAMPLE          RANGE = 500.000 PPM        CO2 =X.XXX   
 
 < TST  TST >  CAL  
 
 
SETUP
 
EXAMPLE 
SAMPLE          
ENTER SETUP PASS: 818
 
 
        
 8        1        8 
 
                       
ENTR
  
 EXIT
 
SETUP X.X 
 
      
PRIMARY SETUP  MENU
 
  
 
 CFG  DAS
  
RNGE
  
PASS  CLK  
MORE
  
   EXIT
 
 
 
DIAG
SIGNAL I  / O   
 
 PREV NEXT 
JUMP 
 
                       ENTR
 
  
EXIT
 
 
 
DIAG  I / O          
Test Signals Displayed Here
 
 
PREV  NEXT
 JUMP 
                      
PRNT  EXIT
 
Use JUMP to go 
directly to a specific 
signal location  
See Appendix A-4 for 
a complete list of 
available SIGNALS 
 
 
 
DIAG   I /  O 
JUMP TO: 12
 
 
 
 1     2                       
 
        ENTR
  EXIT
 
EXAMPLE: 
Enter 12 to Jump to 
12)
 ST_CONC_VALID 
 
 
 
DIAG  I / O         
ST_CONC_VALID = ON
 
 
  PREV NEXT
 
JUMP
 
                ON   
PRNT 
EXIT
 
Exit to return 
 to the  
DIAG
 menu 
SETUP X.X      
SECONDARY SETUP MENU
  
 
 
COMM  VARS  
DIAG   
ALRM
  
      
   EXIT
 
Use NEXT & PREV to 
move between signal 
types. 
Pressing PRNT will send a formatted printout to the serial port and can be captured 
with a computer or other output device. 
EXIT
 returns 
to the main 
SAMPLE
 display  
 
6.13.3. Analog Output Step Test 
This test can be used to check the accuracy and proper operation of the analog outputs.  The test forces 
all four analog output channels to produce signals ranging from 0% to 100% of the full scale range in 20% 
increments.  This test is useful to verify the operation of the data logging/recording devices attached to 
the analyzer.