Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
124 
To configure the analyzer’s three analog outputs, set the electronic signal type of each channel and 
calibrate the outputs.  This consists of: 
  Selecting an output type (voltage or current, if an optional current output driver has been installed) 
and the signal level that matches the input requirements of the recording device attached to the 
channel, See Section 6.13.4.1. 
  Calibrating the output channel.  This can be done automatically or manually for each channel, see 
Section 6.13.4.2 and 6.13.4.3. 
  Adding a bipolar recorder offset to the signal, if required (Section 6.13.4.2.) 
In its standard configuration, the analyzer’s outputs can be set for the following DC voltages.  Each range 
is usable from -5% to + 5% of the nominal range. 
Table 6-17:    Analog Output Voltage Ranges 
RANGE 
MINIMUM OUTPUT 
MAXIMUM OUTPUT 
0-0.1 V 
-5 mV 
+105 mV 
0-1 V 
-0.05 V 
+1.05 V 
0-5 V 
-0.25 V 
+5.25 V 
0-10 V 
-0.5 V 
+10.5 V 
The default offset for all ranges is 0 VDC.