Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
127 
To calibrate the outputs as a group, activate the ANALOG I/O CONFIGURATION MENU (See Section 
6.13.1), then press: 
 
 
 
 
 
 
 
 
DIAG AIO 
AUTO
 
CALIBRATING CONC_OUT_1 
 AUTO
 
CALIBRATING CONC_OUT_2 
 
AUTO CALIBRATING TEST_OUTPUT 
   
 
 
Exit at any time 
to return to the 
main DIAG 
menu 
 
 
 
DIAG  
ANALOG I / O CONFIGURATION 
 
PREV
 
NEXT
 
      
                       ENTR   
EXIT
 
 
 
 
DIAG AIO 
AOUTS CAL IBRATED: 
NO 
 
   < SET
  
SET>
    CAL 
 
 
 
 
   
EXIT
 
If AutoCal has been 
turned off for any 
channel, the message 
for that channel will be 
similar to: 
NOT AUTO CAL 
CONC_OUT_1 
Exit to return to 
the I/O 
configuration 
menu 
STARTING FROM DIAGNOSTIC MENU 
(see Section 6.13.1) 
If any of the channels have 
not been calibrated this 
message will read NO
 
 
 
DIAG AIO      AOUTS CALIBRATED:
          YES
  
 
   < SET
  
SET>    CAL
 
 
  
 
EXIT 
 
To automatically calibrate a single analog channel, activate the ANALOG I/O CONFIGURATION MENU 
(See Section 6.13.1), then press: