Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
125 
The following DC current output limits apply to the current loop modules: 
Table 6-18:   Analog Output Current Loop Range 
RANGE 
MINIMUM OUTPUT 
MAXIMUM OUTPUT 
0-20 mA 
0 mA 
20 mA 
These are the physical limits of the current loop modules, typical 
applications use 2-20 or 4-20 mA for the lower and upper limits.  Please 
specify desired range when ordering this option. 
The default offset for all ranges is 0 mA. 
Pin assignments for the output connector at the rear panel of the instrument are shown in Table 6-19. 
 
ANALOG OUT 
        A1         A2         A3         A4 
    +     -     +     -     +     -     +     -
 
 
 
Table 6-19:   Analog Output Pin Assignments 
PIN 
ANALOG 
OUTPUT 
VOLTAGE 
SIGNAL 
CURRENT 
SIGNAL 
A1 
V Out 
I Out + 
Ground 
I Out - 
A2 
V Out 
I Out + 
Ground 
I Out - 
A3
V Out 
I Out + 
Ground 
I Out - 
A4 
V Out 
Not used 
8 Ground 
Not 
used 
1
 Output A3 is only used when the 
O
2
 sensor option is installed
 
See Figure 3-2 for the location of the analog output connector on the instrument’s rear panel.