Operating Instructions 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
126 
6.13.4.1. Analog Output Signal Type and Range Span Selection 
To select an output signal type (DC Voltage or current) and level for one output channel, activate the 
ANALOG I/O CONFIGURATION MENU (See Section 6.13.1) then press: 
 
 
 
 
DIAG AIO          
CONC_OUT_2 RANGE: 5V 
 
               
SET>
    EDIT 
 
 
 
 
EXIT
 
 
 
 
DIAG 
       
ANALOG I / O CONFIGURATION 
 
PREV
 
NEXT
 
      
          ENTR 
 
 
EXIT
 
 
 
 
DIAG AIO 
AOUTS CALIBRATED: NO  
 
   < SET
  SET>    
CAL
 
 
 
 
 
EXIT
 
FROM ANALOG I/O CONFIGURATION MENU 
DIAG AIO OUTPUT RANGE: 
5V 
 
 
0.1V    1V     5V
    10V 
  CURR     
   
 
ENTR
 
EXIT
 
To set the signal 
level and type of 
the selected 
channel 
 
 
 
DIAG AIO           
CONC_OUT_2:5V, CAL 
 
   < SET  SET>
    EDIT 
 
 
 
 
EXIT
 
Press SET> to select the 
analog output channel to be 
configured. Press EDIT to 
continue 
Pressing ENTR records the new setting 
and returns  to the  previous  menu. 
Pressing EXIT ignores the new setting and 
returns to the previous menu.
DIAG AIO OUTPUT RANGE: 
10V 
 
 
0.1V    1V     5V
  
  10V
 
  CURR     
   
 ENTR  EXIT
 
 
 
6.13.4.2. Analog Output Calibration Mode 
The analog outputs can be calibrated automatically or manually.  In its default mode, the instrument is 
configured for automatic calibration of all channels.  Manual calibration should be used for the 0.1V range 
or in cases where the outputs must be closely matched to the characteristics of the recording device.  
Outputs configured for automatic calibration can be calibrated as a group or individually.  Calibration of 
the analog outputs needs to be carried out on first startup of the analyzer (performed in the factory as part 
of the configuration process) or whenever re-calibration is required.