Troubleshooting 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
225 
seconds, DS5 should flash on and off.  If characters are written to the front panel display but DS5 does 
not flash then the program files have become corrupted.  If after 30 – 60 seconds neither the DS5 is 
flashing or no characters have been written to the front panel display then the CPU is bad and must be 
replaced. 
 
Motherboard 
CPU Status LED 
 
Figure 10-3: 
CPU Status Indicator 
 
10.1.4.2. Sync Demodulator Status LED’s 
Two LEDs located on the Sync/Demod Board and are there to make it obvious that the GFC Wheel is 
spinning and the synchronization signals are present: 
 
Table 10-3:   Sync/Demod Board Status Failure Indications 
LED 
FUNCTION 
FAULT STATUS
INDICATED FAILURE(S) 
D1 
M/R Sensor Status 
 
(Flashes slowly) 
LED is stuck 
ON or OFF 
GFC Wheel is not turning 
M/R Sensor on Opto-Pickup Board failed 
Sync/Demod Board failed 
JP 4 Connector/Wiring faulty 
Failed/Faulty +5 VDC Power Supply (PS1) 
D2 
Segment Sensor 
Status 
 
(Flashes quickly) 
LED is stuck 
ON or OFF 
GFC Wheel is not turning 
Segment Sensor on Opto-Pickup Board failed  
Sync/Demod Board failed  
JP 4 Connector/Wiring faulty 
Failed/Faulty +5 VDC Power Supply (PS1)