Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Troubleshooting 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
227 
Table 10-5:   Relay Board Status LED Failure Indications 
LED FUNCTION 
SIGNAL I/O PARAMETER 
DIAGNOSTIC TECHNIQUE 
ACTIVATED BY 
VIEW RESULT 
D2 
Yellow 
WHEEL 
HEATER 
WHEEL_HEATER WHEEL_TEMP 
Voltage displayed should change.  If not: 
Failed Heater 
Faulty Temperature Sensor 
Failed AC Relay 
Faulty Connectors/Wiring 
D3 
Yellow 
BENCH 
HEATER 
BENCH_HEATER BENCH_TEMP 
Voltage displayed should change.  If not: 
Failed Heater 
Faulty Temperature Sensor 
Failed AC Relay 
Faulty Connectors/Wiring 
D4 
Yellow 
SPARE 
N/A N/A 
N/A 
D5 
Green 
SAMPLE/CAL 
GAS VALVE 
OPTION 
CAL_VALVE N/A 
Sample/Cal Valve should audibly change states.  If 
not: 
Failed Valve 
Failed Relay Drive IC on Relay Board 
Failed Relay Board 
Faulty +12 VDC Supply (PS2) 
Faulty Connectors/Wiring 
D6 
Green 
ZERO/SPAN 
GAS VALVE 
OPTION 
SPAN_VALVE N/A 
Zero/Span Valve should audibly change states.  If 
not: 
Failed Valve 
Failed Relay Drive IC on Relay Board 
Failed Relay Board 
Faulty +12 VDC Supply (PS2) 
Faulty Connectors/Wiring 
D7 
Green 
SHUTOFF 
VALVE OPTION 
SHUTOFF_VALVE N/A 
Shutoff Valve should audibly change states.  If not: 
Failed Valve 
Failed Relay Drive IC on Relay Board 
Failed Relay Board 
Faulty +12 VDC Supply (PS2) 
Faulty Connectors/Wiring 
D8 
Green 
IR SOURCE 
IR_SOURCE CO
2
_MEASURE 
Voltage displayed should change.  If not: 
Failed IR Source  
Faulty +12 VDC Supply (PS2) 
Failed Relay Board 
Failed IR Photo-Detector  
Failed Sync/Demod Board 
Faulty Connectors/Wiring