Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Troubleshooting 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
226 
 
D1 – M/R Sensor Status 
D2 – Segment Sensor Status 
JP4 Connector to Opto-Pickup 
Board 
 
Figure 10-4: 
Sync/Demod Board Status LED Locations 
10.1.4.3. Relay Board Status LEDs 
There are eight LEDs located on the Relay Board.  The most important of which is D1 is the most 
important LED, which indicates the health of the I
2
C bus.  If D1 is blinking the other faults following LEDs 
can be used in conjunction with DIAG menu signal I/O to identify hardware failures of the relays and 
switches on the relay (See Section 6.9.2 and Appendix D).   
Table 10-4:   I2C Status LED Failure Indications 
LED 
FUNCTION 
FAULT STATUS 
INDICATED FAILURE(S) 
D1 
(Red) 
I
2
C bus Health 
(Watchdog 
Circuit) 
Continuously ON  
or 
Continuously OFF 
Failed/Halted CPU 
Faulty Motherboard, or Relay Board 
Faulty Connectors/Wiring to/from Relay Board 
Failed/Faulty +5 VDC Power Supply (PS1) 
 
STATUS LED’s 
DC VOLTAGE TEST 
POINTS 
RELAY PCA 
 
 
Figure 10-5: 
Relay Board Status LEDs