Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
36 
3.7.4. Functional Check 
1.  After the analyzer’s components has warmed up for at least 30 minutes, verify that the software 
properly supports any hardware options that were installed.   
2.  Check to make sure that the analyzer is functioning within allowable operating parameters.  
Appendix C includes a list of test functions viewable from the analyzer’s front panel as well as 
their expected values.  These functions are also useful tools for diagnosing performance 
problems with your analyzer (Section11.1.2).  The enclosed Final Test and Validation Data sheet 
(part number 04307) lists these values before the instrument left the factory. 
To view the current values of these parameters press the following button sequence on the analyzer’s 
front panel.  Remember until the unit has completed its warm up these parameters may not have 
stabilized.   
 
 
RANGE
RANGE1
RANGE2
O
2
 RANGE
STABIL 
MEAS 
REF 
MR RATIO 
PRES 
SAMP FL 
SAMP TEMP 
BENCH TEMP 
WHEEL TEMP
BOX TEMP 
PHT DRIVE 
SLOPE 
OFFSET 
TEST 
TIME
 
 SAMPLE       
RANGE = 500.000 PPM
      CO2 = XXX.X 
 
 
< TST
  
TST >
  CAL 
 
 
 
SETUP 
Only appears instrument is set 
for DUAL or AUTO reporting 
range modes. 
2
 Only appears if O
2
 Sensor 
Option is installed. 
Refer to 
Table 6-2 for 
definitions of 
these test 
functions. 
Toggle <TST TST> buttons 
to scroll through list of 
 
If your network is running a dynamic host configuration protocol (DHCP) software package, the 
Ethernet feature will automatically configure its interface with your LAN. (See Section 6.11.6.2). 
This configuration is useful for quickly getting an instrument up and running on a network. 
However, for permanent Ethernet connections, a static IP address should be used. (See Section 
6.11.6.3).