Справочник Пользователя для Teledyne gfc 7000ta

Скачать
Страница из 280
Getting Started 
 Model GFC7000TA Carbon Dioxide Analyzer 
 
 
Teledyne Analytical Instruments 
38 
2.  Set the expected CO
2
 span gas concentration 
 
 
This sequence causes the 
analyzer to prompt for the 
expected CO
2
  span 
concentration. 
 
 
 
M-P CAL          
CO2 SPAN CONC: 400.000 Conc
 
 
 
 
 0          0          0          
4        5          .0
       
ENTR
   
EXIT
 
  M-P  CAL                  RANGE  = 500.000 PPM            CO2 =X.XXX   
 
 < TST  TST >    ZERO             
CONC
   
EXIT
 
SAMPLE*          RANGE = 500.000 PPM       CO2 =X.XXX  
 
 < TST  TST >  
CAL 
 
 
 
SETUP
 
The CO
2
 span 
concentration values 
automatically default to  
400.0 Conc
.  
 
To change this value to 
the actual concentration of 
the span gas, enter the 
number by pressing the 
button under each digit 
until the expected value 
appears. 
EXIT
 ignores the new setting 
and returns to the previous 
display. 
ENTR 
accepts the new setting 
and returns to the  
previous display.. 
 
NOTE 
For this Initial Calibration it is important to independently verify the precise CO
2
 
Concentration Value of the SPAN gas. 
If the source of the Span Gas is from a Calibrated Bottle, use the exact 
concentration value printed on the bottle.